Features
- 0 to 60 °C (32 to 140 °F) non-freezing process temperature
- 0 to 4 bar (0 to 58 psi) relative process pressure range
- 1 to 12 pH range
- G 1/4" (ISO 228) process connection
- 5 l/h (1.32 gal/h) required sample flow
- Flow alarm
- IP67 protection rated (flow switch)
The E+H Flowfit CYA27 flow assembly is specially designed to accommodate all sensors needed for reliable measurements for the purpose of liquid analysis. The measurement of up to 6 parameters such as disinfection, pH, ORP, conductivity and oxygen is possible. An additional flow measurement enables safe and reliable operation of the measuring point. Integrated flow monitoring ensures that the measurement is always ready for operation. LEDs indicate whether the flow rate is correct and facilitate a quick response in the event of errors.
E+H Flowfit CYA27 flow assembly applications
- Drinking water
- Process watrer
- Salt water
- Swimming pool water